HANNA® instruments Catálogo General v38

PC Connectivity Logged data can be transferred to a Windows compatible PC with the included HI920015 micro USB cable and HI92000 software. Long Battery Life The display of the meter has a battery icon indicator to show the remaining power. The meter uses four 1.5V AA batteries that provide up to 200 hours of battery life. Calibration Error Messages Calibration is successfully performed if the reading is within certain limits. Wrong Buffer — The pH reading is not within range of the selected buffer. Electrode Dirty/Broken alternatively with Buffer Contaminated —The offset of the electrode is not in the accepted range. Check if the electrode is broken or clean it following the Cleaning Procedure at the end of this section. Check the quality of the buffer. If necessary, change the buffer. Wrong or Wrong Old Slope — An inconsistency between new and previous (old) calibration is detected. Calibrate right in the case with custom beaker holders Our custom carrying case features beaker holders for calibration out in the field. Supplied Complete in a Rugged Custom Carrying Case The HI98190 meter, probe, and accessories are supplied in the HI720190 rugged carrying case designed to provide years of use. The inside compartment of the carrying case is thermoformed to securely hold and protect all of the components. HI12963 pH Electrode • Titanium body · Titanium construction provides an unbreakable structure and allows the transfer of heat to the internal temperature sensor for rapid temperature compensation. • Maintenance free, gel-filled electrode · No fill solution required 2 pH 2.68 | www.hannainst.com portable

RkJQdWJsaXNoZXIy MTI3NTM=